【Background】
The product is a proprietary immunology diagnostic kit that is used in effectively diagnosing NPC, a cancer originating in the uppermost region of the throat. The WHO estimates that 80% of NPC cases occur in Southern China and Southeast Asia, threatening more than 400 million people. Our product is the first non-invasive early detection tool available in China, with limited competition. This laboratory-based diagnostic kit is capable of testing 90 patient samples in 1½ hours. The kit provides consistent and repeatable results, and derives a standard unit of measurement that can be used for health-risk monitoring.。
【Technical Principle】
Performing an "indirect" ELISA for determining Rta-IgG antibody concentrations The high-purified Rta antigen is coated on a polystyrene microtiter plate. By adding the blood sample in which the detection antibody Rta-IgG is added, forming a complex with the Rta antigen. The complex can be detected by a secondary antibody (goat-anti-human IgG) conjugated with enzyme HRP. After the final wash step the plate is developed by adding an enzymatic substrate TMB to produce a visible signal, which indicates the quantity of Rta-IgG in the sample with its OD value measured by a plate reader.
【Clinical Application】
It is easy to use, and is capable of running on manual mode or automated platforms. The proprietary technology within NPC diagnostic kit lies in the antigens coated onto the wells, which are highly specific to NPC. These DNA-recombinant antigens work on antigen-antibody binding principles and have been tested on thousands of samples and give clinically-accepted accuracies.
Sensitivity for NPC diagnosis:81%
Specificity for NPC diagnosis: 92%
【Character Features】
The rate with the positive reference material:100%
The rate with the negative reference material:100%
Accuracy(CV%):≤15%
Stability: 2-8℃,6 months
Other Serologic EBV detection kits
- EBV VCA-IgA kit (ELISA)
- EBV EA-IgA kit (ELISA)
- EBV NA-IgA kit (ELISA)
- EBV Zta-IgG kit (ELISA)